How to Optimize Temperature Conditioning and Verification for Industrial Sensor/Transmitter Device Manufacturing – Part One

Sensors Insights by Ken Appel

Introduction

This two-part article describes engineering considerations for specifying thermal environments to optimize production throughput of temperature-compensated sensors and transmitters. Part I covers chamber configurations, while Part 2 will address performance related to heat transfer, uniformity, and controls.

Ensuring measurement quality of devices over their specified operating range, especially industrial sensors, they typically need thermal testing at different stages of the production cycle. It requires that engineering look not only at device size and test specifications, but also consider operations need to conserve floor space and maximize flexibility. Achieving quality goals involves many points of thermal stress including conditioning, verification, calibration, life testing, design qualification, failure analysis, quality checks, and regulatory audits.

Production throughput is an ever-increasing challenge as industrial applications call for higher operating temperatures. Optimizing controlled thermal environments for producing these industrial measurement devices hinges on four areas:

  1. Test being performed
  2. Unit under test (UUT)
  3. Throughput goals
  4. Facility infrastructure

All manufacturing processes are subject to many interrelated variables. The test environment is no different and, as such, we will discuss the above points throughout the article, rather than in sequence.

Understanding the Test Environment

Most process measurement devices, such as low, pressure, level, etc., have at least one thing in common: the need for temperature compensation. As the physical size, thermal mass, and need for stimulation increase with each assembly step, the test environment must also grow.

Typically, no single environment can effectively do it all. For example, a small chamber containing low mass PCBs provides a more efficient enclosure than an oversized volume of space.

Alternatively, subassemblies may need more thermal capacity and volume to meet reliability testing or production throughput. Moreover, the space available in a lab or manufacturing floor offer other challenges to proper chamber sizing. Variables that can affect chamber size and configuration are illustrated in figure 1.

Fig: 1: Four main factors influence the optimal test environment. Configuring a controlled temperature environment depends on how you manage four primary areas: test requirements, devices being tested, throughput, and facility infrastructure.
Fig: 1: Four main factors influence the optimal test environment. Configuring a controlled temperature environment depends on how you manage four primary areas: test requirements, devices being tested, throughput, and facility infrastructure.

Matching Test Environment to Throughput Goals

Your device (circuit, sensor, assembled housing) size, mass, and number of Units Under Test (UUT) will be a factor in determining enclosure size. However, the amount of real estate needed on the plant floor for different phases of production is not necessarily fixed. Thoughtful consideration of configuration options can improve productivity and economize on floor space. Three examples of thermal enclosures configured to optimize throughput are shown in the following sections.

Circuit Conditioning and Pre-Production Example

The electronics for processing sensor signals tend to be small and low mass items. Often, these circuits can be conditioned in a small enclosure driven by a thermal cycler with a large Delta T (difference between thermal source capability and set point temperature). The enclosure can also accommodate multiple circuit boards. (Note: Delta T will be covered in Part 2 of this article.)

For devices sensitive to thermal shock, enclosure designs can disperse air around the circuits to prevent direct exposure to the temperature source, as shown in figure 2 right.

Fig. 2: Air source directly on UUT (left) and directed at sides of enclosure (right).
Fig. 2: Air source directly on UUT (left) and directed at sides of enclosure (right).

Subassembly Testing Example

With the build out of devices – applying transducers and housings – larger thermal enclosures are often required. Large chambers consume a lot of floor space, which typically comes at a premium. Stacking of single-zone chambers with independent controls, as shown in figure 3, is a way to double use of floor space per a given area.

Fig. 3: Stacking thermal chambers<BR>doubles available floor space.
Fig. 3: Stacking thermal chambers
doubles available floor space.

Final System Qualification Example

Finished sensor and transmitter products have different areas of temperature sensitivity. The sensing electronics, for example, may not exceed +85°C, while the seals of a pressure transmitter may need to be tested at +125°C.

Multiple zones (2, 3, 4 or more) provide separately-controlled temperature environments. It permits simultaneous testing of devices at different temperatures. A 3-zone chamber, shown in figure 4, can use the center zone to simulate high heat conditions of pipeline flow. The outer zones can be programmed to meet operating temperature extremes of the instrumented electronics.

Fig. 4: Multiple zones allow simultaneous device testing at different temperatures.
Fig. 4: Multiple zones allow simultaneous device testing at different temperatures.

How User Access Affects Throughput

Plan for user access and how it affects production. Various test setups need different amounts of user involvement and real estate. Consider what physical and visual access to UUTs suit your test needs and available floor space.

What do operators need to effectively and safely set up tests? For example, it may be more convenient to attach cables or measurement connections to the UUT outside of the chamber where there is more room to work. That way loading and unloading of devices is faster.

Adding notches in the chamber, in place of or in addition to ports, provide this capability. During transmitter calibration a different configuration may be needed. For example, you may need to access a pressure source at one side of the chamber. But normal loading of devices may be from another side. Multi-door configurations can optimize user convenience and test productivity.

What about visibility? Windows and lights provide visual access to the UUT. Do you need to economize on floor space in the production or reliability test lab? Chamber stacking and casters are a few ways to increase use of space, as well as flexibility to configure how floor space is used. Figure 5 illustrates these configurations.

Fig. 5: Combination hinged and removable door (left). Multiple hinged doors &ndash; opposing or adjacent sides (right).
Fig. 5: Combination hinged and removable door (left). Multiple hinged doors – opposing or adjacent sides (right).

There are many considerations for optimizing throughput from a chamber configuration standpoint (see figure 6). Users need to make physical connections, load and unload devices, and perform these tasks within the confines of available floor space.

Fig. 6: Partial summary of thermal chamber configurations &ndash; physical and visual access (notches, ports, shelves, windows, lights), stacked chambers, and mobility.
Fig. 6: Partial summary of thermal chamber configurations – physical and visual access (notches, ports, shelves, windows, lights), stacked chambers, and mobility.

Now and What's Coming

The above points are meant to help you understand how different configurations can work to your advantage. The business goals will determine tradeoffs between value of increased production and degree of equipment specialization required. In part two of this article, we'll look at how cooling capacity and temperature control methods affect heat transfer and uniformity performance.

About the Author
Ken Appel, Director of Business Development, inTEST Thermal Solutions, Mansfield, MA, USA. He can be reached at [email protected].

Related Stories
Thermistors As Accurate Temperature Sensors Part 1: Introduction and Methods
This two-part article describes how to use a simple voltage divider circuit with a thermistor to achieve high-accuracy temperature readings over broad measurement ranges. Part one discusses the circuit and various temperature estimation methods.

Thermistors As Accurate Temperature Sensors Part 2: Application and Results
In Part 2 of this two-part article we apply our method to a real-world application and show how, by combining estimation methods with thermistor characterization data, we can achieve high accuracy measurements over a wide temperature range using the simple voltage divider circuit.